Appendix DGlossary
The acronyms and terms used in the tutorial, with the chapter that introduces each.
| Term | Meaning | Chapter |
|---|---|---|
| AAF | Anti-aliasing filter: removes what would fold into the band when sampled | 11, 13 |
| Active electrode | An electrode with a buffer or amplifier at the skin, to lower the impedance its lead presents | 12 |
| AFE | Analog front-end: everything between the electrodes and the converter | 1 |
| Applied part (B, BF, CF) | The IEC 60601 classes of patient connection; CF (cardiac) has the strictest current limits | 4 |
| Bootstrapped switch | A sampling switch whose gate rides on the input, so its on-resistance does not depend on the input | 15 |
| CCIA | Capacitively coupled instrumentation amplifier: gain set by input and feedback capacitors | 7 |
| CDAC | Capacitive digital-to-analog converter, the core of a SAR converter | 17 |
| Chopping | Modulating the signal up to a higher frequency before an amplifier and back after it, to remove offset and 1/f noise | 9 |
| CIC, CoI | Cascaded integrator–comb and cascade-of-integrators decimation filters | 20, 23 |
| CMRR | Common-mode rejection ratio: differential gain over common-mode gain | 4, 7 |
| Coherent sampling | Choosing a test tone that completes a whole number of cycles in the record, so the FFT needs no window | 22 |
| Cole model | A tissue-impedance model with a constant-phase dispersion between a low- and a high-frequency resistance | 28 |
| DNL, INL | Differential and integral nonlinearity: errors of each code step and of the transfer curve, in LSB | 14 |
| DRL, RLD | Driven right leg, right-leg drive: feedback that drives the body’s common mode through a third electrode | 12 |
| DSL | DC servo loop: an integrator that removes the electrode offset from the amplifier | 10 |
| Driven shield, guard | A shield or guard held at the signal’s own voltage, so its capacitance draws no current | 4, 12 |
| ΔΣ (delta-sigma) | An oversampling converter that shapes its quantisation noise out of the band | 19 |
| ECG, EEG, EMG, EOG | Electro-cardio-, encephalo-, myo- and oculography: heart, brain, muscle and eye potentials | 2 |
| ECoG, LFP | Electrocorticography and local field potentials, recorded on or in the brain | 2 |
| EIS, bioZ | Electrical impedance spectroscopy, bioimpedance: measuring tissue impedance with a small current | 28 |
| ENOB | Effective number of bits: (SNDR − 1.76 dB)/6.02 | 13 |
| ExG | Any of the surface biopotentials (ECG, EEG, EMG, EOG) | 2 |
| FoM (Walden, Schreier) | Converter figures of merit: energy per conversion step, and SNDR plus bandwidth over power | 14 |
| gm/ID | Transconductance efficiency, the design variable of the gm/ID method | 5 |
| Incremental ΔΣ | A ΔΣ converter reset before each conversion, so it behaves like a Nyquist converter with high resolution | 19, 20 |
| IQ demodulation | Multiplying by in-phase and quadrature references to recover amplitude and phase | 28 |
| kT/C | The thermal noise frozen on a capacitor when a switch opens | 15 |
| LDO | Low-dropout regulator | 27 |
| LNA | Low-noise amplifier, the first gain stage | 8 |
| Moderate / weak inversion | MOSFET operation near or below threshold, where gm/ID is highest | 5 |
| NEF, PEF | Noise and power efficiency factors: an amplifier’s noise for its current, relative to one bipolar transistor | 6 |
| NTF, STF | Noise and signal transfer functions of a ΔΣ loop | 19 |
| OSR | Oversampling ratio: sample rate over twice the signal bandwidth | 13 |
| OTA | Operational transconductance amplifier | 8 |
| PGA | Programmable-gain amplifier | 11 |
| Pseudo-resistor | Diode-connected MOSFETs far below threshold, used as teraohm resistors | 8 |
| PSRR | Power-supply rejection ratio | 27 |
| RF rectification | Radio-frequency interference turned into a low-frequency error by a nonlinear input stage | 12 |
| SAR | Successive-approximation register converter: a binary search with a DAC and a comparator | 18 |
| SFDR, SNDR, THD | Spurious-free dynamic range, signal to noise and distortion, total harmonic distortion | 14 |
| SQNR | Signal-to-quantisation-noise ratio: 6.02N + 1.76 dB for an ideal N-bit converter | 13 |
| StrongARM | A clocked latch comparator with no static current | 16 |
| Tetrapolar | Impedance measurement with separate current and voltage electrodes | 28 |
| Zcm | Common-mode input impedance of the front-end | 4 |