Appendix BReference Tables

The numbers the chapters design with, collected in one place. The biosignal and electrode values are the ones in labs/python/biosig.py, so the labs use them too.

B.1Biosignals

Table B.1. Amplitude and bandwidth of the biopotentials. Orders of magnitude from the sources of Chapter 2; sources differ by a factor of a few, so treat these as design ranges. The same values are SIGNALS in biosig.py.

Signal Amplitude Bandwidth Typical sample rate
ECG (surface) 0.5–4 mV 0.05–150 Hz 500 S/s
EEG (scalp) 5–300 µV 0.5–100 Hz 256 S/s
EMG (surface) 50 µV–5 mV 10–500 Hz 2 kS/s
EOG 10 µV–3.5 mV 0.05–50 Hz 250 S/s
ECoG 10–500 µV 0.5–500 Hz 2 kS/s
LFP (intracortical) 50 µV–2 mV 0.5–300 Hz 1 kS/s
Spikes (extracellular) 50–500 µV 300 Hz–7 kHz 30 kS/s

B.2Electrodes

Table B.2. Electrode models. A series resistance \(R_\mathrm{s}\) and a parallel \(R_\mathrm{p}\)–\(C_\mathrm{p}\) interface, or a constant-phase element for implanted microelectrodes (Chapter 3). The wet values are the 51 kΩ‖47 nF network of the ECG standards.

Electrode Model |Z| at 50 Hz Half-cell / offset
Wet Ag/AgCl 1 kΩ + 51 kΩ ‖ 47 nF 41.5 kΩ 0.222 V; ±300 mV differential offset to tolerate
Dry 5 kΩ + 1 MΩ ‖ 4.7 nF about 560 kΩ drifts with sweat and pressure
Implanted micro 10 kΩ + constant-phase element, n = 0.85 500 kΩ at 1 kHz charge-injection limits apply

B.3IHP SG13G2 device data

Table B.3. SG13G2 numbers used in this tutorial. From ngspice simulations of the IHP-Open-PDK models at the typical corner and 27 °C (labs/spice/), and from the PDK’s mismatch coefficients.

Quantity Value Chapter
Slope factor n; gm/ID in weak inversion 1.22; 31.7 V⁻¹ 5
Specific current per square, NMOS / PMOS 0.78 / 0.20 µA 5
Thermal noise of a MOSFET at 1 µA 18.5 nV/√Hz 5
Flicker coefficient K, NMOS / PMOS 1.6 × 10⁻¹⁰ / 5.4 × 10⁻¹⁰ V²·µm² 5
Pelgrom AVT, NMOS / PMOS 3.9 / 2.2 mV·µm 5
Pelgrom Aβ, NMOS / PMOS 0.5 / 0.33 %·µm 5
MIM capacitor density; matching; smallest cell 1.5 fF/µm²; 1 %·µm; 1.14 × 1.14 µm² (1.95 fF) 17
Gate capacitance in weak inversion about 2.2 fF/µm² 8
npn13G2 at 1 µA: gm/IC, β, input noise 37.7 V⁻¹, about 680, 15.1 nV/√Hz 5
Gate leakage, 250/4 µm NMOS: thin / thick oxide 3–25 pA / about 10⁻²⁰ A 5, 8
Flicker noise, thick-oxide NMOS about 2.7 × the thin-oxide power at equal area 5, 8
Pseudo-resistor near 0 V: one / four thin-oxide pairs 2 × 10¹¹ / 7.6 × 10¹¹ Ω 8
Pseudo-resistor, one thick-oxide pair, 27 °C 5.6 × 10¹³ Ω 8
Core supply; thick-oxide supply 1.2 V; 3.3 V 30

B.4Patient-current limits

Table B.4. Patient leakage and auxiliary current limits, summarised from IEC 60601-1 [41]. Above 1 kHz the AC limits rise in proportion to frequency, up to 100 kHz, but never above 10 mA. Always work from the edition of the standard that applies to your device (Chapter 4).

Applied part Normal condition Single-fault condition
B, BF (body surface) 100 µA 500 µA
CF (cardiac) 10 µA 50 µA
DC, any type 10 µA —

B.5The Open-BioAFE design in numbers

Table B.5. The main design values, with the chapter that sets each. Table 29.1 compares them with the specification.

Block Values Chapter
LNA chopped CCIA, Cin 20 pF, Cfb 200 fF, gain 40 dB, fch 4 kHz, NMOS input devices of 1000 µm² 8–9
Servo and boosting loops 3.3 V integrator, CDSL 4 pF (noise gain 1.32); input impedance about 200 MΩ 10
Gain plan ECG 100 × 1, EEG 100 × 8, EMG 10 × 8 onto ±0.5 V differential 11
Right-leg drive loop gain 100 at 50 Hz, Ro 150 kΩ, 53° phase margin; CMRR-test residual 0.06 mVpp 12
SAR converter 12-bit asynchronous, 14 cycles at radix 1.8, about 1 fF units, 4 pF per side, Vcm switching, seven low-noise comparator decisions, about 11.7 pJ per conversion, 16 kS/s shared by 8 channels 15–18
ΔΣ converter second-order incremental, M = 1024 at 512 kHz, a1 0.5, a2 0.25, Cs 1.1 pF, 16.5 bits worst case, about 1.9 µW per channel 19–20
Reference off-chip 1 µF reservoir capacitor 17
Budgets noise 0.51 µVrms (wet) / 1.26 µVrms (dry); power 6.8 µW per channel (ECG/EEG), 4.9 µW (EMG); about 1.7 mm² of core 29

B.6Constants

Constant Value
Boltzmann constant k 1.380649 × 10⁻²³ J/K
Elementary charge q 1.602177 × 10⁻¹⁹ C
Thermal voltage UT = kT/q 25.9 mV at 300 K; 26.7 mV at 310 K
4kT 1.66 × 10⁻²⁰ J at 300 K; 1.71 × 10⁻²⁰ J at 310 K (body)
kT/C 64 µVrms on 1 pF at 300 K
Mains 230 V / 50 Hz or 120 V / 60 Hz